What is it?
This article aims to explain the steps to take after failure to import images from PACS to Promap.
Having a failure to import images from PACS can be linked to the type of image not supported by the software or a bad configuration of the PACS.
How to investigate?
- If the serie can be imported but not read, maybe this article will be more helpful.
- Try to load other series.
-
Check the PACS configuration with the help of this article. Especially, go in Maintenance Account > Connectivity and check:
-
"AETitle" of the system and the PACS
-
"Port" of the system and the PACS
- "IP address" of the system and the PACS
-
Do a Ping/Check test
-
- Retrieve the log files following those steps.
- Open a customer request via this form. Report the frequency of the issue, whether if loading other series from the PACS worked, and the result of the Ping/Check test.
How to solve?
Usually, the issue comes from a compatibility issue, or a bad PACS configuration:
-
- On the Trinity system: from Maintenance account > Connectivity, change the settings.
- On the PACS side, please contact directly the IT of the hospital.
The final diagnosis will be made by the Koelis Service team, who will contact you quickly.