Context :
A Central unit (CU) should be changed. How to proceed before the change to save the database and installed it on the new one ?
How to proceed ?
- Turn On the Trinity with the defective central unit
- Launch Maintenance account and select the backup section.
- Plug an USB stick and select this device to export the database. Click on then Save.
- A pop-up window opened to confirm a new repertory creation. The export can take up time, if the patient database was importante.
- On the USB stick, a file named 2602 was created, contained all patient data.
You can now turn OFF the Trinity and follow the instructions linked to replace a CU
After the replacement, follow these instructions to import the patient database:
- Launch Maintenance account and select the Distributor account.
- Distributor password: Koelis3DMapping
A password will be requested to access on this account, communicated during your training
- On backup section, the Restore function was available. Plug your USB stick and select the patient database.
- Click on restore and wait.
The restoration of patient database could take up a long time if the installed base was important.
Once the CU replacement was done, you can now complete the IFUF form to confirm the installation.