There is a bubble inside the head of the probe


    What is it?

    An air bubble appeared inside the head of the probe, impacting the quality of the acquired echographic image.

    How to investigate?

    1. Check if there is any crack on the head of the probe
    2. Check if the dome is still properly sealed to the shield of the probe
    3. Ask for the cleaning process of the probe used by the customer
    4. Open a customer request using this form

    How to solve?

    Regardless of the size of the bubble, the probe has to be replaced under the appropriate commercial conditions.

    Examples:

    • Air bubble on a TR probe:

    bulle_crop

    • Crack on a TR probe, dissociating the dome to the shield and leading to the apparition of an air bubble:
        fissure_crop